13 Measuring
C76IB003EN-E
71
Related topics
Making a Backup of the Instrument Settings: Chapter 17.1
13 Measuring
In this chapter, the filling of samples and the measuring procedure including the
detection of bubbles are described.
13.1 General Sample Settings
Sample name
If you have defined automatic sample name parts (see Chapter 10.3), they will be
added to each sample name after measurement has been finished.
The complete sample name, including automatic sample name parts, may consist
of up to 50 characters.
Measurement types
•
If you select "C (Check)", tap the column "Unit/Settings" and select one of
the defined checks (Air Check and Water Check are predefined).
•
If you select "MM (Multiple Measurements)", tap the column "Unit/Settings"
and enter a value between 2 and 10.
•
If you select "TS (Temperature Scan)", tap the column "Unit/Settings" and
enter the "Start", "Step" and "End" temperature of the temperature scan.
Make sure to set the temperature step interval so that the instrument can
reach the end temperature by a whole number of steps (divide the interval
between start and end temperature by whole numbers only to find
appropriate step settings).
S (Standard)
To perform a standard measurement
C (Check)
To perform one of the predefined checks.
It is only possible to select a check which has
been defined for the selected method before.
MM (Multiple Measurements)
To perform several 2 to 10 measurements of a
single sample automatically.
TS (Temperature Scan)
To perform a temperature scan.