15 Handling the Measurement Data
84
C76IB003EN-E
Related topics
Measuring in Harsh Environments: Appendix C.4
15 Handling the Measurement Data
15.1 Defining the Data Memory Settings
Your instrument can store 1000 result data files with or without camera pictures.
After 900 measurements, the instrument issues a reminder for clearing the
memory. If the memory is full, no measurements are possible and in that case no
more measuring results will be saved, so make sure to free memory space in
time.
To avoid reminders and the necessity of deleting data manually, you can set the
data memory to behave as a circular buffer if you use the instrument in low
security mode.
•
Tap <Menu> and select
"Setup > Expert Settings > Data Memory
Settings"
and activate the check box "Automatically delete old
measurement data".
The oldest 100 measurement results will then be automatically deleted as
soon as the number of 900 measuring results has been exceeded in the data
memory.
15.2 Defining Data Columns for the Data Browser
You can define the kind of data which shall be displayed in the data browser. You
can change these settings any time and add further sample parameters to the
result data files.
To set data for data browser display
1. Tap <Menu> and select
"Setup > Data Browser Settings"
.
2. Tap on the list item that you want to change or on the empty field at the
bottom of the list if you want to add an item.
NOTICE
Never use:
•
highly nonpolar solvents (e.g. toluene, hexane, solvent naphtha)
•
strong acids or bases (e.g. nitric acid, sulfuric acid, hydrochloric acid,
caustic soda)
•
strong mechanical action (steel brush).