17 Service Utilities
C76IB003EN-E
105
TIP
Before starting the update of the user interface firmware, export or print relevant
measurement data (see Chapter 15.6) and create a backup of the instrument
settings (see Chapter 17.1).
TIP
For the new instruction manual, select the respective product page on
www.anton-paar.com and send a download request (under the entry
Manuals
in
the
Downloads
tab).
To update the user interface firmware
1. Load the latest firmware into the root directory of your USB flash drive.
2. Insert the USB flash drive into a USB interface of your instrument.
3. Tap <Menu> and select
"Service > Update > User Interface Firmware"
.
4. Select the location (USB flash drive) and the "file name" of the new firmware.
5. Tap <Update>.
6. Follow the instructions on the screen.
7. Restore the instrument settings (see Chapter 17.2).
To update the module firmware
TIP
A diagnosis message is shown if a too old firmware is installed on one of your
modules. If the auto logon for a user with administrator rights is activated, the
firmware update of the module can be started. Otherwise, the administrator must
log on and then start the firmware update (menu
"Service > Update > Module
Firmware"
).
1. Load the latest firmware into the root directory of your USB flash drive.
2. Insert the USB flash drive into a USB interface of your instrument.
3. Tap <Menu> and select
"Service > Update > Module Firmware"
.
4. Select the location (USB flash drive) and file name of the new firmware or
select "Internal" and tap <Update>.
If you select "Internal", the instrument automatically checks the system. In
case any modules run an older firmware version, the instrument will update it
to the version supported by the user interface firmware. No USB flash drive
is required in that case.
NOTICE
To update the firmware, you need administrator rights with activated auto
logon function (see Chapter 10.4.2).