11 Checking, Adjusting and Calibrating the Instrument
C76IB003EN-E
51
2. Highlight the
"Air Check"
or
"Water Check"
and tap <Edit> to open the two-
step "Check Edit" wizard.
3. Perform the following settings:
•
Enter a name for the check.
•
Use the check box "Must be started every" to define if the check is
obligatory and enter a time interval in days.
4. Tap <Next>.
•
Define the maximum allowed lower and upper deviation.
5. Tap <OK>.
To create a custom check
1. Tap <Menu> and select
"Setup > Expert Settings > Check Management"
to open the check administration list.
2. Tap <New> to open the two-step "Check Edit" wizard.
3. Perform the following settings:
•
Enter a name for the check.
•
Select the method to define detailed method settings for the check, e.g. in
the method "Density".
•
Use the check box "Must be started every" to define if the check is
obligatory and enter a time interval in days.
4. Tap <Next>.
•
Select the quantity.
•
Define the check tolerance entering the lower and upper value.
5. Tap <OK>.
To edit a custom check
1. Tap <Menu> and select
"Setup > Expert Settings > Check Management"
to open the check administration list.
2. Highlight a custom check and tap <Edit> to open the two-step "Check Edit"
wizard.
3. Change the settings as required.