16 Using Special Functions
100
C76IB003EN-E
5. Tap <Next>.
6. Define the input quantity/quantities and tap <Next>.
7. Enter the polynomial coefficients.
If you do not enter a value for a coefficient, the coefficient will be set to the
value zero.
8. Tap <OK>.
To program or edit a user table
If you have a literature table or own experimental data about the density of a
binary mixture at different concentrations, you can program this data into a user
table.
TIP
•
The larger the number of data pairs and the higher the accuracy of the data
pairs, the higher the accuracy of the results that you can obtain with your
user table.
•
Your M series instrument will interpolate, but not extrapolate. This means,
that the data pairs that you enter into the instrument should cover the whole
range of values that you want to measure.
1. Tap <Menu> and select
"Setup > Expert Settings > User Function
Management"
to open the user functions list.
2. Tap <New> to program a new user table or highlight an existing user table
and tap <Edit> to open the four-step "User Function" wizard.
3. Perform the following settings:
•
Enter a name for the user table (up to 50 characters long).
•
Use the drop-down box "Function Type" to select "Table".
•
Use the drop-down box "Physical Quantity" to select the type of quantity
you want to calculate.
•
Use the drop-down box "Unit" to select the output unit.
4. Tap <Next>.
•
Activate/deactivate the check box "Protect function against changes by
other users".
•
Enter a comment to describe this user function.
This comment will not be printed, exported or displayed anywhere but is only
an internal description of your user function.
5. Tap <Next>.
6. Define the input quantity and tap <Next>.
7. Enter the values of the table and tap <OK>.