9 of 10
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Power Supply
1230
Testing resistance of the heater coil
Testing resistance of the heater coil
If you believe that your heater coil is cracked or otherwise damaged and has
caused an open circuit, you can check it by measuring its resistance.
To measure the resistance:
1.
Turn the instrument power off.
2.
Disconnect the oven heater leads (P3, P4) from the AC power board.
3.
Use an ohmmeter to measure resistance at the
connectors
.
Acceptable resistance ranges
(in ohms) are given below. Acceptable
resistances range from the nominal value for a new, cold heater to +5% from
the nominal value.
Note
Resistance goes up approxi3% after heating the coil.
Table 1230-5
Resistances of the Heater Coil
Nominal cold heater resistances
Standard oven
(1600 VA)
Fast-ramp oven
(2250 VA)
120 V
9.07 – 9.52
Ω
n/a
200 V
n/a
17.78 – 18.7
Ω
220 V
n/a
21.51 – 22.6
Ω
230 V
33.06 – 34.71
Ω
23.51 – 24.7
Ω
240 V
n/a
25.60 – 26.9
Ω
n/a =
not available
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......