Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet
210
Replacement procedures
7.
Remove the 1/8-inch Swagelok nut using a 7/16-inch open end wrench.
8.
Use a Torx T-20 screwdriver to loosen the three captive screws that attach
the inlet weldment plate to the top of the inlet carrier.
9.
Pull the inlet up out of the inlet carrier. If necessary, you can also slide
the insulation sleeve off of the bottom of the inlet.
Top insert assembly replacement
1.
The top insert assembly is replaced as one assembly if required.
For EPC inlets:
a.
Disconnect the pneumatics block from the front of the EPC module
(one Torx T-10 screw). Be careful not to lose the three O-rings from
the pneumatics module.
b.
Follow the split vent flow line back from the pneumatics block and
disconnect the line from the split vent trap.
c.
Remove the carrier and septum purge lines from the left side of the
GC.
For non-EPC inlets:
a.
Unscrew the top insert assembly (large inlet nut and flow lines on
the top of the inlet) that has the carrier gas and septum purge lines
plumbed to it. Raise this assembly up and away from the inlet.
b.
Follow the split vent flow line back from the pneumatics block
attached to the top insert assembly and disconnect the line from the
split vent trap.
c.
Remove the carrier and septum purge lines from under the plastic
tabs on the left side of the GC.
2.
Use a 7/16-inch wrench to unscrew the split vent flow line nut from the
Swagelok fitting at the top of the inlet.
3.
Use a Torx T-20 screwdriver to loosen the three captive screws that attach
the inlet weldment plate to the top of the inlet carrier.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......