5 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Faults
850
Fault 214—Front detector flame out
If the electrometer is out of specification, the detector will never reach a ready
state.
❐
Check that the cable that connects the detector to the board is connected
properly.
❐
If the boards are connected properly but the electrometer is still out of
specification, the electrometer or detector board is broken.
Fault 214—Front detector flame out
Fault 215—Back detector flame out
Status message: Front det flame out or Back det flame out
This message appears when the FID is not able to ignite or if the flame goes
out during a run. During the ignition process or the run, the detector will try
to ignite the flame twice; if both attempts fail, the hydrogen, air, and ignitor
will shut off, and the error message will appear. The detector will be in a not
ready state.
Check the following:
❐
Make sure the hydrogen and air are turned on and that the flow rates are
high enough for the flame to ignite.
❐
Use an electronic leak detector to search for and correct leaks around
the detector column fitting.
❐
Check that you are using the correct jet for your column.
❐
Change the Lit Offset to 0.5 (the default value).
Fault 216—Front TCD filament open
Fault 217—Back TCD filament open
Status message: F TCD filament open or B TCD filament open
The TCD filament bridge voltage indicates that the filament resistance is too
high. The resistance may be too high because the filament is broken or worn
thin from use, or the wires from the TCD are not connected on the detector
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......