Solvent Vapor Exit Accessory
Leak testing
8 of 10
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
1.
Replace the septum of the Cool On-column inlet.
2.
Using a 1/16-inch (SS) union and the appropriate graphite/Vespel
ferrules, connect the detector end (250
µ
m) of the tri-column assembly
and the transfer line (320
µ
m) together.
3.
From the 6890 GC keyboard, set the inlet pressure to 50 psi and wait
approximately 15 seconds for equilibration.
Note
If pressure can not be reached, either a very large leak is present or the supply
pressure is not high enough (at least 10 psi greater than desired pressure).
4.
Cap the septum purge vent of the Cool On-column flow module using a
solid (no- hole) 1/8-inch Vespel ferrule (part no. 0100-1372) and a 1/8-
inch Swagelok nut (part no. 5180-4103). As an alternative a 1/8-inch
Swagelok cap can be used.
5.
From the GC keyboard, turn the inlet pressure “OFF”.
6.
Note the “Actual” reading on the display and monitor for 10 minutes. You
can use the stopwatch feature of the 6890 GC to monitor the time by
pressing [
Time
] and then [
Enter
] to start timing.
7.
If there is less than 1.5 psi pressure loss, consider the system to be leak
tight.
8.
If pressure loss is much greater than 1.5 psi, there is a leak that must be
found and corrected.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......