Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Faults
850
Fault 223—Oven thermal shutdown
❐
If the oven sensor is operating correctly and the error persists, the main
board must be replaced.
❐
See
Fault 221—Thermal shutdown (status message)
.
Fault 223—Oven thermal shutdown
Status message: Oven temp too cool
The oven temperature reading indicates the oven is less than its minimum
allowable temperature. This indicates a stuck open cryo valve or a shorted
oven sensor.
Power to the small zones is disabled. To clear the error message, turn off all
the heated zones and turn the GC off and then on again.
❐
See
Fault 221—Thermal shutdown (status message).
Fault 224—Oven thermal shutdown
Status message: Oven temp sensor
The oven temperature reading reports the oven’s temperature is lower than
expected, which indicates a shorted oven sensor. Power will be turned off for
all the heated zones.
❐
The sensor must be replaced.
❐
See
Fault 221—Thermal shutdown (status message).
Fault 225—Front detector thermal shutdown
Status message: F det temp too hot
The front detector’s temperature is higher than the maximum allowable
temperature (detector type maximum temperature plus an additional 25
°
C).
This indicates a defective heater drive on the mainboard or a defective sensor.
Power is disabled to all the heated zones. This fault can only be cleared by
turning off all the heated zones and turning the instrument off and on again.
❐
The problem could be caused by a defective sensor or defective heater
driver on the mainboard.
❐
See
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......