7 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning Codes
830
Warning 124—Host communications: abnormal break
Warning 124—Host communications: abnormal break
Status message: Comm abnormal break
There is an interruption in communication between the 6890 GC and a PC or
workstation. This could be caused by a LAN disturbance or a bad cable
connection. Data may be lost
❐
If you are connected to the LAN, check to see if it is down or there is a
problem with your connection to it.
❐
Check the cable connection between the GC and the computer or
workstation.
Warning 125—Sampler communications: data overrun
Status message: Sampler data overrun
The 6890 GC received information from the 7673 ALS quicker than it could
process it. The data may be lost.
❐
Verify that the sampler settings are adjusted correctly.
Warning 126—Sampler communications: data error
Status message: Sampler data error
The 6890 GC received bad information from the 7673 ALS. For example, there
may be a framing or parity error.
❐
Verify that the sampler settings are adjusted correctly.
Warning 127—Sampler communications: abnormal break
Status message: Sampler abnormal com
There is an interruption in communication between the 6890 GC and the
7673 ALS. This could be caused by a bad cable connection. Data may be lost.
❐
Check the cable connections on the GC and the sampler.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......