6 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
•
On/off valves mounted on the outlet fitting of the two-stage regulator are
not essential but are very useful. Be sure the valves have stainless-steel,
packless diaphragms.
•
FID, FPD, and NPD detectors require a dedicated air supply. Operation
may be affected by pressure pulses in air lines shared with other devices.
•
Flow- and pressure-controlling devices require at least 10 psi (138 kPa)
pressure differential across them to operate properly. Source pressures
and capacities must be high enough to ensure this.
•
Auxiliary pressure regulators should be located close to the GC inlet
fittings. This insures that the supply pressure is measured at the
instrument rather than at the source; pressure at the source may be
different if the gas supply lines are long or narrow.
Supply tubing for carrier and detector gases
Caution
Do not use methylene chloride or other halogenated solvent to clean tubing
that will be used with an electron capture detector. They will cause elevated
baselines and detector noise until they are completely flushed out of the
system.
Gases should be supplied to the instrument only through preconditioned
copper tubing (part no. 5180-4196). Do not use ordinary copper tubing—it
contains oils and contaminants.
Caution
Do not use plastic tubing to supply detector and inlet gases to the GC. It is
permeable to oxygen and other contaminants that can damage columns and
detectors, and can melt if near hot exhaust or components.
The tubing diameter depends upon the distance between the supply gas and
the GC and the total flow rate for the particular gas. One-eighth-inch tubing
is adequate when the supply line is less than 15 feet (4.6 m) long.
Use larger diameter tubing (1/4-inch) for distances greater then 15 feet (4.6
m) or when multiple instruments are connected to the same source. You
should also use larger diameter tubing if high demand is anticipated (for
example, air for an FID).
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......