Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Not Ready Messages
820
Detector not ready
Front det igniting
Back det igniting
Status message: Front det igniting or Back det igniting
The GC is not ready while the FID or FPD is going through the flame ignition
sequence. The messages will clear if the detector is turned off.
If the FID is unable to ignite, there may be a problem, and the detector may
eventually shutdown.
Front det adjusting
Back det adjusting
Status message: Front det adjusting or Back det adjusting
The GC will be not ready because the NPD or ECD is adjusting its baseline to
reach the offset (NPD) or output (ECD) setpoint. The ECD’s adjustment is
usually complete in 30 seconds. The NPD may take longer.
It is possible that the NPD will be unable to reach the setpoint if there is
contamination in the system (for example, if the gas is not pure enough or
the bead is damp) or if the bead is worn out. If the NPD cannot reach the
setpoint, you will not receive an error message; the GC will simply not become
ready.
Turn the detector off to clear the message.
Front det equib time
Back det equib time
Status message: Front det equib time or Back det equib time
The NPD has completed adjusting the offset and is now waiting for the value
to remain at the setpoint for the equilibration time. Equilibration time can be
changed from the Detector control menu.
The NPD may not be able to equilibrate if the system is contaminated or the
bead is worn out. In addition, changes in the room temperature could prevent
equilibration. The GC will become ready if the detector is turned off.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......