340
Electron Capture Detector (ECD)
Diagnostics
18 of 24
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
4.
If the ECD frequency indicates contamination (
≥
100) check for the
following:
•
Contaminated carrier gas trap(s) and or supply—replace carrier gas
supply tank and any traps on the carrier supply line.
•
Insufficient column conditioning—fully condition the column.
•
Contaminated detector—bake out the detector.
•
Column, inlet and/or septum bleed—clean the inlet/replace the
septum with a conditioned, low bleed septum.
•
Leaks—perform leak tests on both the inlet and detector systems.
•
Anode current leakage—make sure the anode contacts are clean.
Make sure the anode nut is tight.
Leak test
Note
Once you have determined that the flow system components upstream from
the detector (gas supply tubing, inlet, column fittings) are leak free, perform
the following ECD detector leak test.
1.
With the GC on and operating normally, set the oven, detector, and inlet
temperatures to ambient.
2.
Turn off the ECD and then turn off the inlet pressure.
3.
Turn off the anode and makeup gas flows.
4.
Cap the ECD exhaust vent with a vent plug (part no. 5060-9055).
5.
Set carrier gas pressure at the inlet corresponding to the ECD to 15 psi
(103 kPa).
6.
Wait until the system reaches the setpoint pressure and then turn off the
pressure and monitor the actual pressure value for at least 10 minutes.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......