5 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
External Connectors, 6890A and 6890 Plus
1210
APG remote start/stop
Signal descriptions
Prepare (low)
—
Request to prepare for analysis. Receiver is any module
performing pre-analysis activities.
Start (low)
—
Request to start run/timetable. Receiver is any module
performing runtime-controlled activities.
Shutdown (low)
—
System has serious problem. Receiver is any module
capable to reduce safety risk.
Power On (high)
—
All modules connected to system are switched on. Receiver
is any module relying on operation of others.
Ready (high)
—
System is ready for next analysis. Receiver is any sequence
controller.
Stop (low)
—
Request to reach system ready state as soon as possible (for
example, stop run, abort or finish, and stop injection). Receiver is any module
performing runtime-controlled activities.
Start Request (low)
—
Request to start injection cycle (for example, by a
Start
key on any module). Receiver is the automatic liquid sampler.
APG remote control
Remote control allows easy connection between single instruments or
systems to ensure coordinated analysis with simple coupling requirements.
To provide maximum safety within a distributed analysis system, one line is
dedicated to SHUTDOWN the system’s critical parts in case any module
detects a serious problem.
To detect whether all participating modules are switched on or properly
powered, one line is defined to summarize the POWER ON state of all
connected modules.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......