520
Gases
Cryogenic cooling requirements
12 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Using liquid nitrogen
WARNING
Liquid nitrogen is a hazard because of the extremely low temperatures and
high pressures that may occur in improperly designed supply systems.
Liquid nitrogen can present an asphyxiant hazard if vaporizing nitrogen
displaces oxygen in the air. Consult local suppliers for safety precautions and
design information.
Liquid nitrogen is supplied in insulated Dewar tanks. The correct type for
cooling purposes is a
low-pressure
Dewar equipped with a dip tube—to deliver
liquid rather than gas—and a safety relief valve to prevent pressure build-up.
The relief valve is set by the supplier at 20 to 25 psi.
WARNING
If liquid nitrogen is trapped between a closed tank valve and the cryo valve
on the GC, tremendous pressure will develop and may cause an explosion.
For this reason, keep the delivery valve on the tank open so that the entire
system is protected by the pressure relief valve.
To move or replace a tank, close the delivery valve and carefully disconnect
the line at either end to let residual nitrogen escape.
Additional requirements for the liquid N
2
system include:
•
Nitrogen must be provided to the GC as a liquid at 20 to 30 psi.
•
The supply tubing for liquid N
2
must be
insulated
. Foam tubing used for
refrigeration and air-conditioning lines is suitable for insulation. Since
pressures are low,
insulated
copper tubing is adequate.
•
The liquid nitrogen tank should be close (only 5 to 10 feet) to the GC to
insure that liquid, not gas, is supplied to the inlet.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......