Nitrogen Phosphorus Detector (NPD)
320
Correcting NPD hardware problems
Correcting NPD hardware problems
No detector response to injected sample
•
A solvent peak has extinguished the hydrogen/air plasma. Increase the
bead voltage. Run the detector at a higher offset (for example, 40 to 50
pA), or use makeup gas at a flow rate of 5 mL/min.
•
Check that hydrogen is flowing to the detector. The flow rate should be
between 1.0 and 5.5 mL/min.
•
The bead is not activated. Look through the vent hole on the detector lid
to see if the bead is glowing orange. If not, check that there is enough
current reaching the bead. Check the detector background signal. Reduce
the bead voltage to zero to establish a reference level, then look for a
sudden sharp increase in output as the bead voltage increases, which
indicates that plasma ignition occurred. If 4 V are being supplied to the
bead but it is not igniting, the bead is probably burned out. Replace the
bead.
•
The bead power cable is bad.
•
A contaminated upper ceramic insulator will cause a high offset (2 to 15
pA or more) when the bead voltage is off. This directly affects sensitivity.
Replace the ceramic insulator. Minimize the problem by running the detec-
tor hot (320 to 335°C).
No baseline; output signal exceeds 8 million
•
The electrometer ribbon cable is not attached to the PC board properly.
Be sure to turn the GC off before reattaching the cable! If the signal does
not drop to a normal level (<3 pA), replace the electrometer.
•
The collector is shorted to the detector housing. Check the insulators.
Baseline level is 0.0
•
Broken electrometer
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......