5 of 10
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Solvent Vapor Exit Accessory
260
Replacement procedures
7.
Make a loop (or loops) in the excess column protruding from the valve
and move it to an unobstructive position.
•
Coil it in a loose coil (approximately 6 inch diameter).
•
Secure the coil with heat resistant tape or equivalent.
Replacing the tri-column assembly
1.
Cool the inlet and detector.
2.
Turn inlet and detector gases off.
3.
Disconnect the Tri-column assembly from the COC inlet, detector and
(SS) union and remove the column assembly from the GC oven.
4.
Install the new SVE Tri-column assembly onto a column hanger and hang
the assembly inside the GC oven. Position the column assembly so that
the end of the retention gap (530
µ
m) is located under the COC inlet and
the end of the analytical column (250
µ
m) is under the detector.
5.
Connect the retention gap to the COC inlet using a graphite ferrule and
a column nut.
6.
If using a MSD, connect the analytical column (HP-5MS) to the MSD using
a column nut (part no. 05988-20066) and a graphite/Vespel ferrule
(part no.5062-3508). If using any other type of detector, connect the
analytical column (HP-5MS) to the GC detector using a column nut
(part no. 5181-8830) and a graphite/Vespel ferrule (part no. 5062-9527).
7.
Remove the stainless ferrules from the union. Use a graphite/Vespel
ferrule to connect the 0.5 m × 320 mm transfer line from the quartz Y-
splitter of the Tri-column assembly to the unused end of the stainless
union located inside the GC oven.
8.
Check all of the connections for leaks using an electronic leak detector.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......