830
Warning Codes
Warning 128—Front inlet sensor auto zero calib failed
8 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning 128—Front inlet sensor auto zero calib failed
Warning 129—Back inlet sensor auto zero calib failed
Status message: F inl flow cal fail or B inl flow cal fail
An attempted auto calibration of the flow sensor offset for the front or back
inlet exceeded the allowable calibration range. The previous calibration
setpoints remain in memory.
Warning 130—Aux 1 and front inlet on same cryo valve
Warning 131—Aux 2 and back inlet on same cryo valve
Status message: Aux 1 cryo disabled or Aux 2 cryo disabled
The same cryo valve is specified for and inlet and an auxiliary zone. The aux
cryo is declared not installed.
❐
Check the hardware to see where the valve is really installed. Fix the
configuration accordingly.
Warning 132—Col 1 Init time changed
Warning 133—Col 2 Init time changed
Status message: None
While using a Volatiles Interface inlet, the Sample End time is set greater than
the oven Init time. The Init time has been changed to the Sample End time to
avoid the conflict.
Warning 134—Front inlet title mismatch
Warning 135—Back inlet title mismatch
The title for the inlet does not match the title stored in its pneumatics module.
No action is taken. If the inlet is a different type from that stored in the
pneumatics module, a hardware error will be declared.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......