25 of 26
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Micro-cell Electron Capture Detector (
µ
-ECD)
341
Maintaining a
µ
-ECD detector
Maintaining a
µ
-ECD detector
µ
-ECD bakeout (thermal cleaning)
If your
µ
-ECD baseline is noisy or the display frequency is too high (i.e.,
≥
1000), you should perform a thermal cleaning (also called a “bakeout”) of
the detector. Before performing a bakeout, verify that the carrier supply gas
and flow system are leak- and contaminant-free.
Caution
Detector disassembly and/or cleaning procedures other than thermal should
be performed only by personnel trained and licensed appropriately to handle
radioactive materials. Trace amounts of radioactive
63
Ni may be removed
during these other procedures, causing possible hazardous exposure to
β
- and
x-radiation (bremsstrahlung).
WARNING
To prevent possible hazardous contamination of the area with radioactive
material, the detector exhaust vent must always be connected to a fume hood,
or otherwise vented in compliance with the latest revision of Title 10, CFR,
Part 20, or with state regulations with which the Nuclear Regulatory
Commission has entered into an agreement (USA only). For other countries,
consult with the appropriate agency for equivalent requirements.
1.
Record the
µ
-ECD “Output” value from GC display. If the number is equal
to or greater than 1000, you should continue with this procedure.
2.
Remove the column from the detector.
3.
Cap the bottom of the makeup gas adapter with a blank column ferrule
and column nut.
4.
Set the makeup gas flow rate to 60 mL/min. Set the detector temperature
between 350 and 375°C.
5.
Set the oven temperature to 250°C.
6.
Allow thermal cleaning to continue for several hours, and then cool the
system to normal operating temperatures.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......