1230
Power Supply
Setting the instrument power configuration
2 of 10
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Line voltage configuration plug
There is a different line voltage configuration plug on the AC power board for
each power configuration. Each configuration uses three or five jumper wires,
each connecting to two different pins on the plug. The 120 VAC configuration
uses five jumper wires and all other configurations use three jumper wires.
Table 1230-2
Line Voltage Configuration Plugs (P8 on AC board) by Power Option
and Diagram of Pinouts from the Top of the Plug
Ceramic fuses
The two ceramic oven heater fuses on the AC power board (F1, F2) will be
one of two types, depending on the line voltage:
•
One for the 120 V power option
•
One for all other power options.
These fuses should always be
replaced as a pair.
Power option
Fuse rating
120 V
20A/250 V (Type F)
200 V-240 V
15A/250 V (Type F)
Jumper locations for power configuration plug
120 VAC
200 VAC
208/220
VAC
230 VAC
240 VAC
Transformer
3
↔
13
2
↔
13
3
↔
13
5
↔
13
3
↔
13
6
↔
15
6
↔
12
6
↔
12
6
↔
9
6
↔
9
9
↔
14
Oven fan
1
↔
10
1
↔
4
1
↔
4
1
↔
4
1
↔
4
4
↔
8
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...