Agilent 6890 Gas Chromatograph Service Manual
Gases
520
The gas plumbing
Be generous when cutting tubing for local supply lines—a coil of flexible tubing
between the supply and the instrument lets you move the GC without moving
the gas supply. Take this extra length into account when choosing the tubing
diameter.
Two-stage pressure regulators
To eliminate pressure surges, use a two-stage regulator with each gas tank.
Stainless steel, diaphragm-type regulators are recommended.
Figure 520-2
Two-stage pressure regulator
The type of regulator you use depends upon gas type and supplier. The
Chemical Analysis Consumables and Accessories
catalog contains
information to help you identify the correct regulator, as determined by the
Compressed Gas Association (CGA). Agilent Technologies offers pressure-
regulator kits that contain all the materials needed to install regulators
properly.
Pressure regulator-gas supply tubing connections
The pipe-thread connection between the pressure regulator outlet and the
fitting to which you connect the gas tubing must be sealed with Teflon tape.
Instrument grade
Teflon tape (part no. 0460-1266), from which volatiles have
been removed, is recommended for all fittings. Do not use
pipe dope
to seal
the threads; it contains volatile materials that will contaminate the tubing.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......