Flame Photometric Detector (FPD)
350
Replacement procedures
Replacing the transfer line fused silica liner
Occasionally the transfer line fused silica liner between the column and FPD
module must be inspected, cleaned, and/or replaced.
1.
Turn off power to the gas chromatograph and disconnect the main power
cord. Remove the detector covers.
2.
Allow time for heated zones to cool to safe temperatures.
3.
Inside the oven, remove the column to the FPD.
Caution
Always turn the electrometer or the main power off before removing the PMT
housing to avoid destroying the tube.
4.
Remove the photomultiplier tube assembly—or assemblies—from the
detector module; also remove the filter(s). Set them in a safe place. Keep
the open end of the PMT covered whenever practical to avoid light damage
to the tube.
5.
Locate the ignitor cable attached to the side of the detector. Trace the
cable back to the printed circuit board and disconnect it there.
6.
Remove the exhaust tubing and the sheet metal cover—on the single
wavelength detector, it is held by two screws on the top and two at the
bottom; on the dual wavelength detector, it is held by two screws at the
top.
7.
Remove the four screws that attach the detector to the top of the oven
(one at each corner). Remove the detector from the GC.
8.
Loosen the transfer line nut. Remove the two screws that secure the
U-clamp to the detector frame. Remove the U-clamp and the attached
parts from the bottom of the detector.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......