Agilent 6890 Gas Chromatograph Service Manual
Pneumatics Control Module
270
Leak testing the PCM
Leak testing the PCM
Use an electronic leak detector capable of detecting the gas being used. Liquid
leak detectors are not recommended, especially in areas where cleanliness is
very important. If using leak detection fluid, immediately rinse the fluid off
to remove the soapy film.
WARNING
To avoid a potential shock hazard when using liquid detection fluid, be careful
not to spill leak solution on electrical leads, especially the detector heater
leads.
WARNING
Be careful! The oven and interface may be hot enough to cause burns.
Occasionally, to locate small leaks, you will need to isolate the PCM module
from the inlet weldment and leak test the PCM separately.
1.
On the keyboard, turn off pressure to the inlet being tested.
Press
[Front Inlet]
or
[Back Inlet]
, scroll to the Pressure field
and press
[
Off
]
.
2.
Use a Torx T-10 screwdriver to remove the screw in the plumbing block
on the front of the PCM. Remove the plumbing block from the PCM, being
careful not to lose the O-rings between the block and the module.
3.
Replace the inlet’s plumbing block with the leak test block (part no.
G1530-20660) from the leak test kit (part no. G1530-60960). Make sure
you install O-rings (if needed) between the block and the PCM module to
create a seal.
The leak test block is a special fitting that plumbs the carrier gas coming
out of the module directly back into the septum purge and split vent flow
paths on the module. It allows the carrier gas and septum purge line to
function normally as if an inlet were present.
4.
Perform the normal leak test as described previously in this section. With
the inlet removed, the internal volume is quite small and a pressure loss
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......