Agilent 6890 Gas Chromatograph Service Manual
220
Purged/Packed Inlet
The septum purged/packed column inlet contains a mass flow controller to
control the flow stream into the inlet and a forward pressure regulator to
control flow out the septum purge vent. The inlet can be used for packed or
capillary columns. Based on the column you have configured, the inlet can
be operated in either a mass flow controlled mode or a pressure controlled
mode.
Theory of operation
At injection, liquid sample is introduced with a syringe into the hot inlet where
it is flash vaporized in the liner. For both manual and EPC inlets, the flow
paths are the same; the flow is divided into two streams.
•
The major portion of the stream
enters the inlet insert and is transferred
into the column.
•
The remaining small portion of the flow stream
is vented out the purge
vent flow line.
EPC purged packed inlets can be either flow controlled or pressure controlled
depending on whether a capillary column’s dimensions are configured within
the instrument. If configured with a capillary column dimension, the column
is pressure controlled and flows are calculated values.
EPC inlet
Carrier gas flow enters into the top of the inlet through a flow-limiting frit
and then a proportional valve (PV1) in the carrier gas feed line. This valve is
either controlled by the flow sensor (flow-controlled mode) or the pressure
sensor (pressure-controlled mode) also in the carrier gas feed line. A fraction
is split off to purge the septum. 1.0 mL/min flows out the purge vent and the
rest of the stream enters the inlet insert and flows through the column.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......