5 of 6
Jun 2001
Valves
Agilent 6890 Gas Chromatograph Service Manual
Valves
1110
Liquid sample valves
Liquid sample valves
Agilent Technologies offers 4-port LSVs with 0.2, 0.5, or 1 µL internal loops.
These valves are designed for liquefied gases under pressure such as ethane,
propane, butane, LNG, etc. They are not intended for nonvolatile liquids (at
room conditions) where a concealed leak could allow an accumulation or pool
of liquid to form that may present a significant fire hazard.
All liquid sample valves have 1/16-inch fittings. The 0.5 µL size is available in
low- (1000 psig limit) and high-pressure (5000 psig) versions; the other sizes
are low-pressure (1000 psig) only.
An adjustable restrictor may be used on the sample outlet line to maintain
internal sample pressure to keep a compressed gas liquefied.
Temperature ranges
1/16-inch Teflon rotor valves
0 to 200
°
C
Adjustable restrictor valves
Ambient to 225
°
C
Caution
The life of an LSV is shortened if used outside its specified pressure and
temperature ranges. Highly dangerous leaks can occur if the valve box
temperature ever exceeds the specified temperature limits.
Adjustable restrictor valve
The adjustable restrictor supplied with a liquid sample valve is designed for
temperatures up to 225
°
C; it is NOT compatible with high-temperature valves.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......