Agilent 6890 Gas Chromatograph Service Manual
Gases
520
Cryogenic cooling requirements
Cryogenic cooling requirements
Cryogenic cooling allows you to cool the oven below ambient temperature. A
solenoid valve introduces liquid coolant, either carbon dioxide (CO
2
) or
nitrogen (N
2
), to cool the oven to the desired temperature.
CO
2
and N
2
require different hardware. You must replace the entire valve
assembly if you want to change coolants. The liquid CO
2
valve kit is
part no. G1565-65510 and the liquid N
2
kit is part no. G1566-65517.
Choosing a coolant
When selecting a coolant, consider these points:
•
The lowest temperature you need to reach
•
How frequently you will use cryogenic cooling
•
The availability and price of coolant
•
The size of the tanks in relation to the size of the laboratory
•
Liquid N
2
cools reliably to –80°C
•
Liquid CO
2
cools reliably to –40°C
CO
2
is the choice for
infrequent
cryogenic cooling because it does not
evaporate and is less expensive than N
2
. However, a tank of CO
2
contains
much less coolant than a tank of N
2
and more CO
2
is used for the same amount
of cooling.
Although liquid N
2
evaporates from the tank regardless of frequency of use,
N
2
tanks contain more coolant than do CO
2
tanks and therefore may be better
for frequent use.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......