340
Electron Capture Detector (ECD)
Replacement procedures
10 of 24
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Replacing the makeup gas adapter
After removing the ECD detector from the GC, you can further disassemble
it to replace the makeup gas adapter. The makeup gas adapter consists of a
line from the detector pneumatics manifold that carries makeup gas to a
weldment that screws into the bottom of the ECD detector. From there, the
makeup gas sweeps past the end of the column and carries the column effluent
into the ECD cell.
1.
Remove the detector as described in the Replacing the entire detector/
detector cell procedure in this section.
2.
Remove the Torx T-20 screw holding the pneumatics block(s) to the
detector manifold.
Note
There are two pneumatics blocks on the EPC version of the ECD pneumatics
manifold. The outside block is the anode purge gas line and the inside block
is the makeup gas line.
3.
Slide the makeup gas adapter up and out of the GC.
4.
When reinstalling the makeup gas adapter, ensure the following:
•
Approximately 6 inches of the makeup gas line should reside in the
oven after installation.
•
The makeup gas line should be bent into a coil that loops around the
bottom of the detector weldment and makeup gas adapter.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......