240
Programmed Temperature Vaporization Inlet (PTV)
Leak testing the PTV module only
42 of 46
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Leak testing the PTV module only
Occasionally, to locate small leaks, you will need to isolate the EPC module
from the inlet weldment and leak test the EPC module separately.
1.
On the keyboard, turn off pressure to the inlet being tested.
Press
[Front Inlet]
or
[Back Inlet]
, scroll to the Pressure field and
press
[Off]
.
2.
Use a Torx T-10 screwdriver to remove the screw in the plumbing block
on the front of the module. Remove the plumbing block from the EPC
module, being careful not to lose the O-rings between the block and the
module.
3.
Replace the inlet’s plumbing block with the leak test block (part no.
G1530-20660) from the leak test kit (part no. G1530-60960). Make sure
you install O-rings (if needed) between the block and the EPC module to
create a seal.
The leak test block is a special fitting that plumbs the carrier gas coming
out of the module directly back into the septum purge and split vent flow
paths on the module. It allows the carrier gas, septum purge, and split
vent line to function normally as if an inlet were present.
4.
Perform the normal leak test for the EPC inlet as described previously in
this section. With the inlet removed, the internal volume is quite small
and a pressure loss of 0.5 psi or less in
5
minutes time is considered to
be leak free.
5.
If there is a leak, you will probably need to remove the flow manifold to
try to isolate the leak as described in the following Locating leaks on the
flow manifold procedure.
6.
If there is an
increase
in pressure, see the following Forward pressure
valve leaks procedure.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......