3.3. General
3.3.1. Interface
Control I/O
Input: open collector Output: photo coupler
DIN connector 8 pin
Standard serial output
Output for communication with A&D peripheral devices
(20 mA current loop) DIN connector 7 pin
RS-232C
(Standard equipment)
D-sub 9 pin male
RS-232C/RS-485 (Option)
It is possible to change the current loop to the RS-232C
(ch2) or RS-485.
RS-232C AD4410-04
RS-485 AD4410-03
3.3.2. Measuring functions
Zero adjustment
•
Sets the gross weight to zero by pressing the ZERO key.
•
Selection of disable or enable for the operation when unstable.
•
The zero value is stored in the non-volatile memory.
•
Zero adjustable range
Can be set optionally in the range of 1 to 100% of the
weighing capacity.
Zero tracking
•
Tracks the weight drift around the zero point to maintain zero.
•
Zero tracking time
0.0 to 5.0 seconds (Can be set optionally within the range.)
•
Zero tracking band
0.0 to 9.9 d (Can be set optionally within the range.)
Tare
•
Sets the net weight to zero by pressing the TARE key.
•
Selection of disable or enable for the operation when
unstable or negative.
•
The tare weight is stored in the non-volatile memory.
•
Tare range: Gross weight
≤
Weighing capacity
Stability detection
•
Turns ON the stabilization indicator when the variables of
the weight values per sampling are within the set band in
the set time.
•
Stability detection time
0.0 to 9.9 seconds (Can be set optionally within the range.)
•
Stability detection band
0 to 9 d (Can be set optionally within the range.)
Digital filter
Cutoff frequency range (-3dB) : 0.07 to 11 Hz
Near zero detection function
Detects whether there is a load or not and outputs the result
as near zero.
Comparison function
HI, OK, or LO is output by comparing a value to an upper or
lower limit value.
Hold function
This is displayed by holding the weighing value.
Normal hold / Peak hold / Averaging hold