10.3.2. All data initialization
Step 1 Remove the CAL key cover and press the CAL key.
Cal
is displayed to indicate that the indicator enters the calibration mode.
Step 2 Press the ENTER key to display
Cal5et
. The indicator enters the mode of
calibration with an actual load. To go back to the weighing mode, press the ESC
key.
Step 3 Press the
∧
or
∨
key to select the all data initialization mode (
iniallLL
) and
press the ENTER key.
Step 4 All the status LEDs blink to prompt a confirmation. To initialize, press and hold
the ENTER key for three seconds or more.
After initialization, the indicator is reset and all the display segments are ON. The
indicator then enters the weighing mode.
To exit from this mode without performing initialization, press the ESC key.