7.7. Linearity-related functions (LnrFnc)
Confirm and change linearity settings.
To use this function, select “
lnrf
” in the same way as calibration-related functions are
selected.
Function No.
Parameter or
Setting range
Function Description
Default
value
Llnrf01
0
to
5
Number of
input points
Number of points where linear input was
done.
The linear-zero input is included as one
point.
Digital linearization is not performed
when the set value is between
0
and
2
.
0
Llnrf02
-7.00000
to
7.00000
Linear-zero
Voltage during linear-zero input
Units in mV/V.
0.00000
Llnrf03
0
to
999999
Linear 1
mass value
The mass value during linear 1 input
The decimal point position depends on
Calf02
.
0
Llnrf04
0.00000
to
9.99999
Linear 1
span
The span voltage from linear-zero during
linear 1 input
Units in mV/V.
0.00000
Llnrf05
0
to
999999
Linear 2
mass value
The mass value during linear 2 input
The decimal point position depends on
Calf02
.
0
Llnrf06
0.00000
to
9.99999
Linear 2
span
The span voltage from linearity zero
during linearity 2 input
Units in mV/V.
0.00000
Llnrf07
0
to
999999
Linear 3
mass value
The mass value during linear 3 input
The decimal point position depends on
Calf02
.
0
Llnrf08
0.00000
to
9.99999
Linear 3
span
The span voltage from linearity zero
during linearity 3 input
Units in mV/V.
0.00000
Llnrf09
0
to
999999
Linear 4
mass value
The mass value during linear 4 input
The decimal point position depends on
Calf02
.
0
Llnrf10
0.00000
to
9.99999
Linear 4
span
The span voltage from linearity zero
during linearity 4 input
Units in mV/V.
0.00000