Page
13
AD-4410
5.2. Type of load cell cables
The load cell cable must have high insulation and shield performance.
Use a shield cable that is made with a material with high insulation resistance, such as
Teflon or polyethylene.
For best results, we recommend using A&D's load cell cables.
AX-KO162-5M to 100M (5m to 100m)
Cable diameter…
φ
9mm
Cross-sectional area of the conducting wire…0.5mm
2
6 wire cable equipped
5.3. Connecting the shield wire
Connect the shield wire of the load cell cable to only the shield terminal (7th pin of the load
cell connector) of the AD-4410.
As a general rule, do not ground between the load cell and AD-4410. This is because when
there are multiple ground points, a ground loop can be created along the wiring route, and
interference such as high-frequency noise becomes more likely.
5.4. Load cell connection
Connecting the shield wire
As a general rule, only connect the load cell cable's shield wire to the weighing indicator's
shield terminal. This is because when there are multiple ground points, it becomes easier
for noise to occur as a result of a ground loop. However, when there are higher priorities
than noise prevention, such as explosion prevention systems or static electricity release
routes, this rule does not necessarily need to be followed.
Connecting the signal wire
AD-4410 has two connection configurations: the 6-wire configuration and the 4-wire
configuration.
We recommend connecting using the 6-wire configuration for optimum accuracy and
stability.