AD-4410
<
>
∧
∨
keys: Press to correct the value of the
calibration weight used.
CAL key
: Press to check the current value
(the gross weight) and
press again to display the
calibration weight value.
ESC key
: Press three times to go back to
the weighing mode without
performing span calibration.
Step 6 Place a calibration weight on the load cell. Wait
for the stabilization indicator to turn ON and
press the ENTER key. - - - - - - is displayed
for approximately two seconds.
Step 7
Calend
is displayed. To re-adjust the span,
press the F key.
Step 8 Press the ESC key.
Cal5et
is displayed and
the calibration data is written into the FRAM.
Step 9 The current state is the same as that of step 2. To
go back to the weighing mode, press the ESC key
again.
NOTE: The blinking decimal point indicates that the current value is not the weight
value.
If
C err
X is displayed, an error has occurred. Refer to “7.8. Calibration
Errors” to take some measures.