NOTE: Record the setting values of
Calf17
,
18
and
19
in the “Setting List” at
the end of the manual to prepare against a failure.
By changing the parameters of
Calf17
,
18
and
19
, “zero calibration”
and “span calibration” can be adjusted optionally. (Digital span accuracy
approx. 1/5000: The accuracy varies depending on the load cell output
accuracy and the conditions of calibration.)
Except for an emergency, perform calibration with an actual load.
Calf20
0
to
1
Output when
the gross weight
is negative
Output by the manual print command when
the gross weight is negative.
0
: Enables output.
1
: Disables output.
0
Calf21
0
to
1
Communication
restriction
Restriction on network-related
communication such as changing station
numbers or ID numbers and outputting by
auto printing
0
: Disables restriction.
1
: Enables restriction.
0
Calf22
1
to
2
Header 2
Second header of serial output
1
: GS/NT/TR
2
: G_/N_/T_ ( _ is space, ASCII 20h)
1
Calf23
1
to
2
Number of digits for
unit in serial output
data
1
: 2 digits
2
: 3 digits
1
Calf24
1
to
6
Scale interval of
dual range
Minimum division (d) of second range
1
: 1
2
: 2
1
CALF17
Input
voltage
CALF19
CALF18
Displayed
weight
0
0