AD-4410
Page
32
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf01
0
to
5
Unit
Unit of the weight value
0
: None
1
: g
2
: kg
3
: t
4
: N
5
: kN
2
Calf02
0
to
5
Decimal
point
position
Decimal point position of the weight value
0
: None 123456
1
: 10
1
12345.6
2
: 10
2
1234.56
3
: 10
3
123.456
4
: 10
4
12.3456
5
: 10
5
1.23456
0
Calf03
1
to
6
Minimum
division
Minimum division (d) of the weight value
1
: 1
2
: 2
3
: 5
4
: 10
5
: 20
6
: 50
1
Calf04
1
to
999999
Weighing
capacity
Weighing capacity of the weighing instrument
Weighing is possible up to the value of this
setting plus 8 d. If the value exceeds this,
overflow will occur and will not be
displayed.
The decimal point position is the same as
the setting of
Calf02
.
70000
Calf05
0
to
100
Zero
adjustment
range
Range to enable zero adjustment by the
ZERO key
Expressed as a percentage of the weighing
capacity with the calibration zero point as
the center.
For example, if this is set to 2, the value in
the range of
±
2% of the weighing capacity
with the calibration zero point at the center
will be to zero.
When a power-ON zero is performed, the
initial zero point will be the center.
2