Page
37
AD-4410
3
: 5
4
: 10
5
: 20
6
: 50
If this is set to less than the scale interval of
Range 1, the instrument will revert to single
range mode. The scale interval for Range 1 is
Calf03
.
Calf25
0
to
999999
Dual range
border
When the absolute weighing value (gross
weight or net weight) is over the set range, it
will become Range 2.
The weighing capacity for Range 2 is
Calf04
.
The decimal point position is linked to
Calf02
.
0
Calf26
9.75000
to
9.85000
Gravity
acceleration of
place of
calibration
Gravity acceleration of the place of
calibration.
Units in m/s
2
9.80000
Calf27
9.75000
to
9.85000
Gravity
acceleration of
place of use
Gravity acceleration of the place where the
scale is being used.
Units in m/s
2
9.80000
Calf28
0
to
1
Disable
averaging hold
0
: Enabled
1
:
Disabled
0
NOTE: To receive a Measuring instruments directive (MID) approval, network-
related communication must be restricted. Confirm that “
1
: Enables
restriction” is selected for
Calf21
(Communication restriction).