AD-4410
Page
44
7
: 1.4 Hz
8
: 1.0 Hz
9
: 0.7 Hz
10
: 0.5 Hz
11
: 0.33 Hz
12
: 0.25 Hz
13
: 0.17 Hz
14
: 0.13 Hz
15
: 0.10 Hz
16
: 0.07 Hz
fncf07
1
to
3
Hold
1
: Normal hold
2
: Peak hold
3
: Averaging hold
1
fncf08
-999999
to
999999
Set value of
near-zero
The reference value for near-zero
Decimal point position is linked to
Calf02
.
10
fncf09
1
to
2
Mass to be
compared
with
near-zero
Item to be compared with near-zero
1
: Gross weight
2
: Net weight
1
Ffncf10
-999999
to
999999
Set value of
upper limit
Reference value for the upper limit
The decimal point position is linked to
Calf02
.
10
Ffncf11
-999999
to
999999
Set value of
lower limit
Reference value for the lower limit
The decimal point position is linked to
Calf02
.
-10
Ffncf12
1
to
2
Mass to be
compared
with upper
and lower
limit
Item to be compared with the upper and
lower limit
1
: Gross weight
2
: Net weight
1
Ffncf13
1
to
2
Output logic
of upper
and lower
limit
Logic used when the result of the
comparison with the upper and lower limit is
output
1
: Positive logic
2
: Negative logic
1