7.2. Calibration With an Actual Load (
Cal5et
)
Calibration is performed using a calibration weight
Before performing calibration for the first time, set the unit, decimal point position,
minimum division and weighing capacity in the calibration-related function mode.
NOTE: To avoid drift caused by changes in temperature, warm up the indicator for
ten minutes or more before performing calibration with an actual load.
Step 1 Press the ON/OFF key to turn the indicator ON
(weighing mode) if it is in the OFF mode.
Remove the CAL key cover and press the CAL
key.
CalL
is displayed to indicate that the indicator
has entered the calibration mode.
Step 2 Press the ENTER key to display
Cal5et
.
The indicator enters the mode of calibration with
an actual load. To go back to the weighing mode,
press the ESC key.
Zero calibration
Step 3 Press the ENTER key to display
Cal 0
.
If zero calibration is not to be performed, press
the F key and go to step 5.
To monitor the current weight value, press the
CAL key. Press the CAL key again to display
CalL 0
.
Step 4 With nothing placed on the load cell, wait for the
stabilization indicator to turn ON and press the
ENTER key. - - - - - - is displayed for
approximately two seconds.
If span calibration is not to be performed, press
the ESC key twice to return to the weighing
mode.
Span calibration
Step 5 Press the ENTER key when
Cal5pn
is
displayed.
The calibration weight value (the weighing
capacity currently set) is displayed and the
lowest value number blinks.