AD-4410
Page
54
9.2.5. Bit structure
Data 7 bits
Data 8 bits
9.2.6. Data transfer mode
Four data transfer modes are available; stream, auto printing, manual printing or jet
stream.
Stream mode
Transmits data in sync with the display update rate. If
the indicator can not catch up with the update rate
because of the baud rate, the transmission is paused
until the next updating.
The transmitted data is the same as what is being
displayed. Data that is not displayed is never
transmitted.
Auto printing mode
Transmits data one time once a weight value of 5 d or
greater has stabilized.
For a further transmission, the weight value must fall
below 5 d and then stabilize again above that value.
Manual printing mode
Transmits data when the F key, assigned as the
manual print command, is pressed.
Jet stream
Transmits data in sync with the sampling rate. If the
indicator can not catch up with the sampling because
of the baud rate, the transmission is paused until the
next sampling.
Functions related to the data transfer mode
z
Set
Cl f04
,
r51f06
or
r52f06
(ID number) to add an ID number to standard serial
output.
Example:
z
Set
Calf20
(disables negative output)
to disable the print command when gross
weight is negative.
z
Set
Calf22
(header 2)
to change header 2 to G, N or T.
S T , G S , + 0 1 2 3 . 4 5 k g
CR LF
0 9
@
ID number
Stop bit
Stop bit
Start bit
Data bit
Data bit
Parity bit