When the AD-4410 does not operate properly, write the required items in the table below
and contact your local A&D dealer.
Item
User’s usage circumstances
Model number, rated,
measurement value etc.
Description
Load cell connection
method
□
4-wire configuration
□
6-wire configuration
When using the 4-wire configuration,
it is necessary to connect a jumper
between EXC+ and SEN+, and
between EXC- and SIG-.
Model number of load cell
used
Rated capacity of the load
cell
[Unit ]
Rated output of the load
cell
[mV/V]
Allowable overload of the
load cell
[%]
Number of load cells used
Use of summing box
Length of the extension
cable
[m]
Length from the indicator to the
summing box etc.
Initial load on the weighing
instrument
[Unit ]
Minimum
division
of the
weighing instrument
[Unit ]
Write all digits including decimal
figures.
Example: 0.002kg
Weighing capacity of the
weighing instrument
[Unit ]
Write all digits including decimal
figures.
Example: 10.000kg
Load cell output during
initial load (when no load is
applied)
[mV/V]
From -0.1mV/V to rated
sensitivity value
(With the first load)
Load cell output when
loaded to capacity (Or
when a mass of choice is
loaded)
Load cell output at
Load [Units ]
[mV/V]
When loaded to capacity, the
output value of the initial load +
the rated output value of the load
cell.
(Must be within allowable overload)
Measurement point
Measurement contents Measurement
result
EXC+
①
SEN+
②
Decrease in voltage on the
EXC+ side of the load cell
cable
[mV]
EXC+
①
EXC-
④
Load cell excitation voltage
[V]
SEN-
③
EXC-
④
Decrease in voltage on the
EXC- side of the load cell
cable
[mV]
SIG-
⑥
EXC-
④
Load cell center point voltage
[V]
SIG+
⑤
SIG-
⑥
Load cell output voltage
[mV]