Page
35
AD-4410
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf13
1
to
3
Exceeding
negative
gross weight
To judge when the negative gross weight
is exceeded.
A/D negative exceeded or,
1
: Gross weight < -999999
2
: Gross weight < negative weighing capacity
3
: Gross weight < -19 d
1
Calf14
1
to
2
Exceeding
negative net
weight
To judge when the negative net weight is
exceeded.
Gross weight negative exceeded or,
1
: Net weight < -999999
2
: Net weight < negative weighing capacity
1
Calf15
0
to
1
Clear the zero
value
Select whether or not to clear the zero
value.
0
: Disables.
1
: Enables.
1
Calf16
0
to
1
Zero setting
when power is
turned on
Select whether or not to perform zero
setting when power is turned on.
0
: Disables.
1
: Enables.
0
Calf17
-7.00000
to
7.00000
Input voltage
at zero
Input voltage from a load cell at zero.
Unit: mV/V
This value is determined in zero
calibration during the calibration with an
actual load.
0.00000
Calf18
0.00001
to
9.99999
Input voltage
at span
Input voltage from a load cell at span
Unit in mV/V
This value and the value of
Calf19
are
determined in span calibration during the
calibration with an actual load.
3.20000
Calf19
1
to
999999
Calibration
weight value
corresponding
to Input
voltage at
span
The calibration weight value corresponding
to the Input voltage at span of
Calf18
.
When performing digital span,
Calf17
,
18
and
19
are required.
The decimal point position is the same as
the setting of
Calf02
.
32000
Содержание AD-4410