Page
65
AD-4410
10.3. Initialization Mode
Initialization restores the contents of the FRAM to the default values.
Three types of initialization are available as shown below.
Initialization mode
Display
Description
RAM initialization
ini ra
Initializes the contents of the RAM only.
The zero and tare values will be restored
to 0.
General functions
initialization
inifnc
Initializes the contents of the general
functions stored in the FRAM.
All data
initialization
iniall
Initializes all the contents in the FRAM.
The calibration-related data is also initialized.
So, calibration is required before using
the indicator again.
10.3.1. RAM initialization / General functions initialization
Step 1 While pressing and holding the ENTER key, press the F key.
fnc
is displayed to indicate that the indicator will enter the general function
mode.
To go back to the weighing mode, press the ESC key.
Step 2 While pressing and holding the ZERO key, press the ENTER key.
1
Chc
is displayed to indicate that the indicator will enter the check mode.
Step 3 Press the
∧
or
∨
key to select the initialization mode (
ini
) and press the
ENTER key.
Step 4 Press the
∧
or
∨
key to select the item to be initialized and press the ENTER
key.
Step 5 All the status LEDs blink to prompt a confirmation. To initialize, press and hold
the ENTER key for three seconds or more.
After initialization, the indicator is reset and all the display segments are ON. And
the indicator enters the weighing mode.
To exit from this mode without performing initialization, press the ESC key.