AD-4410
Page
34
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf08
0.0
to
9.9
Stability
detection
time
Performs stability detection using this setting
in combination with the setting of the stability
detection band.
When 0.0, stability detection will not be
performed. (Stable all the time)
Unit: second in 0.1 increments
1.0
Calf09
0
to
9
Stability
detection
band
Performs stability detection using this setting
in combination with the setting of the stability
detection time.
When 0, stability detection will not be
performed. (Stable all the time)
Unit: d (minimum division)
2
Calf10
0
to
1
Tare and
zero
adjustment
when
unstable
Tare and zero adjustment when the weight
value is unstable.
0
: Disables both functions.
1
: Enables both functions.
1
Calf11
0
to
1
Tare when
the gross
weight is
negative
Tare when the gross weight is negative.
0
: Disables tare.
1
: Enables tare.
1
Calf12
0
to
1
Output
when
overflow
and
unstable
Standard serial output when the weight value
overflows and is unstable.
0
: Disables output.
1
: Enables output.
1
Stability detection outputs the STABLE signal when changes in
the weight value are within a certain range during a certain time.
Weight
value
CALF09
STABLE
signal
CALF08
CALF08
CALF09