Page
77
AD-4410
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
Calf13
1
to
3
To judge when the negative gross weight is
exceeded.
A/D negative exceeded or,
1
: Gross weight < -999999
2
: Gross weight < negative weighing capacity
3
: Gross weight < -19 d
1
Calf14
1
to
2
To judge when the negative net weight is
exceeded.
Gross weight negative exceeded or,
1
: Net weight < -999999
2
: Net weight < negative weighing capacity
1
Calf15
0
to
1
Select whether or not to clear the zero value.
0
: Disables.
1
: Enables.
1
Calf16
0
to
1
Select whether or not to perform zero upon
power-ON.
0
: Disables.
1
: Enables.
0
Calf17
-7.00000
to
7.00000
Input voltage from a load cell at zero.
X.XXXXX mV/V
0.00000
Calf18
0.00001
to
9.99999
Input voltage from a load cell at span
X.XXXXX mV/V
3.20000
Calf19
1
to
999999
The value of the calibration weight
corresponding to the span input voltage
32000
Calf20
0
to
1
Select whether or not to output when the
gross weight is negative.
0
: Enables output.
1
: Disables output.
0
Calf21
0
to
1
Select whether or not to restrict
network-related communication.
0
: Disables restriction.
1
: Enables restriction.
0
Calf22
1
to
2
Second header for serial output
1
: GS / NT / TR
2
: G_ / N_ / T_ ( _ is space, ASCII 20h)
1
Calf23
1
to
2
Number of unit digits for serial output
1
: 2 digits
2
: 3 digits
1
Calf24
1
to
6
Scale interval of dual range
1
: 1
2
: 2
3
: 5
4
: 10
5
: 20
6
: 50
1