71
AD-4410
11.2. Hold
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
LHldf01
0.00
to
9.99
Averaging time
Units in seconds.
Averaging is not done for 0.00.
0.00
LHldf02
0.00
to
9.99
Start wait time
Units in seconds.
0.00
LHldf03
0
to
2
Automatic start conditions
0
: Do not use automatic start.
1
: Above near-zero range and stable
2
: Above near-zero range
0
LHldf04
0
to
1
Release when control input falls
0
: Do not release
1
: Release
1
LHldf05
0.00
to
9.99
Release after time limit
Units in seconds.
0.00 will not cause a release.
0.00
LHldf06
0
to
999999
Release due to fluctuation
The decimal point position depends on
Calf02
.
0 does not cause a release.
0
LHldf07
0
to
1
Release within near-zero range
0
: Cancellation has no function
1
: Cancellation has function
0