AD-4410
Key switches
No. Name
Description
9 CAL
•
The key to enter the calibration mode.
•
The key is concealed by a sealable cover.
To avoid unintentional operations, keep the cover attached
except when entering the calibration mode.
10
•
The key to turn the indicator ON (Weighing mode) and OFF (OFF
mode). To turn the indicator OFF (OFF mode), press and hold this
key.
•
When OFF, all of the displays and external output/input devices
are turned OFF, but a decimal point located at the right side of
the main display remains ON.
•
The key functions as an ESC key when the numerical values are
being entered.
11
•
The key to proceed to the selected mode.
•
Stores the data when pressed after a setting is changed.
12
•
The function for this key can be selected in the general
functions, from the list below:
No function
Manual print command
Hold
Alternate switch
Momentary switch
Clear the tare weight
Clear the zero value
•
The key decreases the value of the blinking digit by one when
the numerical values are being entered.
13
•
The key to switch between the gross weight and the net weight.
•
The key increases the value of the blinking digit by one when the
numerical values are being entered.
14
•
The key to perform tare.
•
The tare conditions are set in the calibration-related functions.
•
The key shifts the position of the blinking digit to the right when
the numerical values are being entered.
•
In the OFF mode, pressing the ON/OFF key while holding the
TARE key will clear the zero value and the tare weight.
15
•
The key to zero the current display.
•
The zero conditions are set in the calibration-related functions.
•
The key shifts the position of the blinking digit to the left when
the numerical values are being entered.
16 Seal
•
Cover be sealed using a wire seal.
ON/OFF
ESC
ENTER
Net/B/G
NET GROSS
TARE
ZERO