AD-4410
Page
42
Step 6 Press the ESC key. The function number disappears and the indicator returns to
the state of step 3.
Press the ESC key to store the setting values in the FRAM and go back to the
weighing mode.
NOTE: The blinking decimal point indicates that the current value is not the weight
value.
If a value exceeding the settable range is entered,
err dt
is displayed and
the input is canceled.
8.2. Adjusting the digital filter
Adjust the digital filter using
fncf05
(digital filter).
The available settable range for the cutoff frequency is from 11 Hz to 0.07 Hz.
The cut off frequency is the frequency where vibrations begin to decline.
1. If the weighing value is unstable, set the cutoff frequency lower.
2. To make the response faster, set the cutoff frequency higher.
Cut off frequency is high.
Response rate is fast.
Weak against disturbances.
Cut off frequency is low.
Response rate is slow.
Strong against disturbances.
* It is possible to make adjustments while watching the effects of the digital filter with your own
eyes.
By pressing the CAL key while setting
fncf05
(digital filter), it is possible to check the
weight displayed.
ZERO key Set the display to zero.
∧
Key Increase stability.
∨
Key Increase speed.
CAL key Return to the value setting display.