6.8.1.2. Adjusting tilt eucentricity
Before you can start imaging or milling, it might be necessary to adjust tilt eucentricity. By adjust-
ing the eucentricity, the specimen surface is moved into the tilting plane of the super-eucentric
stage. That is why the image does not shift out of the screen when the stage is tilted.
CAUTION
Danger of damaging objective lens or specimen if the specimen is too close to the objec-
tive lens.
Since the eucentricity is adjusted by using the M-axis, the working distance will be
changed during the eucentricity setup.
Ensure the working distance is 10 mm or more.
Procedure
1
Open the
Panel Configuration Bar
.
2
Double-click
FIB Daily Adjust
.
The
FIB Daily Adjust
panel opens.
3
To show the crosshairs, activate the
Cross-
hairs
checkbox .
4
Center a characteristic feature in the middle of
the screen (i.e. in the middle of the crosshairs).
5
Click
Eucentric Axis
.
6
Click
Start
.
7
Follow the instructions in the wizard.
To re-centre the feature, use the Centre feature
function (
<Ctrl+Tab>
) or change X/Y.
To change the tilt degree of the stage:
a
In the
SEM Control
panel, select the
Stage
tab.
b
In the
Go To T
(ilt) field, enter the required
degree.
Summary of Contents for Crossbeam 340
Page 1: ...Crossbeam 340 Crossbeam workstation Instruction Manual ...
Page 24: ...OQ çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF OK p ÑÉíó p ÑÉíó ÉèìáéãÉåí ...
Page 60: ...SM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF RK fåëí ää íáçå ...
Page 120: ...NOM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 126: ...NOS çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NNK ÄÄêÉîá íáçåë ...
Page 130: ...NPM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...
Page 133: ......