lÄí~áåáåÖ=~=Ñáêëí=áã~ÖÉ
6.4.3. Locating the specimen
Procedure
1
In the TV mode (ChamberScope), look at the image area of the user interface to observe the
sample holder and the specimen inside the chamber.
CAUTION
Risk of damaging the objective lens and/or your specimen.
Ensure not to hit the objective lens while driving the stage. Watch the moving stage in TV
mode.
2
To open the
Stage Navigaton
bar, on the
menu, select
Stage/Navigation
.
The
Stage Navigation
shows specimen
chamber schematics. The upper schematic
shows a lateral view of the specimen stage.
The lower schematic shows a plan view of the
stage with the different stubs.
3
Press
Additional
Settings
.
The
Stage Nav Settings
dialog appears.
4
Press
Show Gallery
.
The
Sample Holder Gallery
dialog appears.
5
Double-click your stub.
6
Close the
Sample Holder Gallery
dialog and
the
Stage Nav Settings
dialog.
Summary of Contents for Crossbeam 340
Page 1: ...Crossbeam 340 Crossbeam workstation Instruction Manual ...
Page 24: ...OQ çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF OK p ÑÉíó p ÑÉíó ÉèìáéãÉåí ...
Page 60: ...SM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF RK fåëí ää íáçå ...
Page 120: ...NOM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF UK qêçìÄäÉëÜççíáåÖ mçïÉê ÅáêÅìáí ...
Page 126: ...NOS çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NNK ÄÄêÉîá íáçåë ...
Page 130: ...NPM çÑ NPQ fåëíêìÅíáçå j åì ä êçëëÄÉ ã PQM ÉåMOE loF NPK aÉÅä ê íáçå çÑ ÅçåÑçêãáíó ...
Page 133: ......